Weight test head
The weight test head performs an automated repeatability test
by placing a small test weight on given tare weights. The
repeatability of the system is determined based on 10
consecutive measurements of the test weight. If your
instrument fails the tests, inform a METTLER TOLEDO service
technician.
Liquid dosing head
Liquid dosing heads are used for automated liquid dosing.
They are used together with a pump and bottle. The liquid
dosing head contains an RFID tag which stores information
about the dosing head and the used substance.

3.5 Dosing head RFID tag

Each powder and liquid dosing head is equipped with an
integrated RFID tag (1) that stores and exchanges data with
the instrument.
Various data is stored in the RFID tag of the dosing heads,
such as the substance name, lot ID, filling date, expiry date,
1
1
etc. It also includes customized data fields.
This data can be edited on the terminal and should be set
before using a new dosing head such that the data is
available for reports and labels.
For powder dosing heads, the counter for the remaining
powder dosing cycles is based on the Dose limit of the
dosing head, which can be defined by the customer. With
every dosing cycle started, the counter is decreased by 1. If the counter drops to zero, consider replacing the
dosing head. If the vial of the old dosing head still contains a considerable amount of powder, you may
remove the vial from the old dosing head and screw it onto the new dosing head. Copy the user data and the
powder content value from the old dosing head to the new dosing head.
The quantity of powder added to the dosing head can be stored in the RFID tag during the filling procedure. The
RFID information is updated after each dosing operation such that the powder quantity in the dosing head is up
to date. If the remaining quantity of powder is insufficient for the next dosing cycle, a warning message will
appear on the terminal.
For further information, consult the Reference Manual (RM) of your XPR balance.
u
www.mt.com/XPR-analytical-RM

4 Installation and Putting into Operation

This instrument must be installed by a METTLER TOLEDO service technician.
Q3 Dosing Module
11