Connect a tube (1) to the exhaust air outlet (2) to collect
the contaminated air into a safe container (3).
Note
Outer tube diameter: 6 mm
The exhaust air tube and the container are not provided
by METTLER TOLEDO.

5 Operation

For further information, consult the Reference Manual (RM) of your XPR balance.

5.1 Installing and removing the dosing head

When the dosing module is in its uppermost position (Home position), the dosing head is unlocked. That
means that it can be released by pressing the release button of the holder. During dosing, the dosing head is
locked and cannot be removed until it returns to its uppermost position.
Installing the dosing head
The dosing module is in its uppermost position.
1 Slide the dosing head (1) onto the dosing head holder
(2) until it comes to a stop.
2 Slightly press down the dosing head until it properly sits
in the dosing head holder (2).
The dosing head is locked in the holder.
The dosing head is ready to dose.
18
3
2
1
u
www.mt.com/XPR-analytical-RM
1
2
Q3 Dosing Module