Spo Setup (Optional) - Gima VITAL PRO-SNET Manual Del Usuario

Ocultar thumbs Ver también para VITAL PRO-SNET:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

4.9.2 SpO
Setup (Optional)
2
Figure 4.12A SpO
Setup Screen
2
The SpO
setup screen is as shown in figure 4.12A or 4.12B, please refer to the monitor in your hand.
2
Screen Description:
² "SpO
ALM": SpO
2
"Hi/Lo": high and low alarm limit for SpO
² "PR ALM": pulse rate alarm switch..This is set "ON" and the user cannot adjust it.
"Hi/Lo": high and low alarm limit for PR.
²
"PI% display": PI display switch. "√" means PI display is enabled, "×" means PI display is disabled.
²
"Response speed": "Normal", "Steady" and "Fast" for optional. Note: this setting item is optional, please refer
to the monitor you purchased.
²
AvgTime (only for Masimo SpO
8, 10, 12, 14 and 16 seconds.For the 2 and 4 second settings the averaging times may range from 2-4 and 4-6
seconds, respectively. The default is 8. The stability of the SpO2 and PR readings over time is affected by the
averaging mode being used.The longer the averaging time,the more stable the readings tend to become.This is
Figure 4.12B SpO
Figure 4.12C SpO
Setup Screen (with Nellcor SpO
2
alarm switch; This is set "ON" and the user cannot adjust it.
2
is configured): The signal averaging time of the oximetry can be set to: 2, 4,
2
Setup Screen (with Masimo SpO
2
2.
30
configuration)
2
configuration)
2

Publicidad

Tabla de contenido
loading

Tabla de contenido