NIBP Setup Screen Description:
SYS ALM, DIA ALM, MAP ALM are set "ON" and the user cannot adjust them.
² "SYS ALM": systolic pressure alarm switch.
"SYS Hi/Lo": high and low alarm limit for systolic pressure
² "DIA ALM": diastolic pressure alarm switch.
"DIA Hi/Lo": high and low alarm limit for diastolic pressure.
² "MAP ALM": mean arterial pressure alarm switch.
"MAP Hi/Lo": high and low alarm limit for mean arterial pressure.
² Initial pressure: Initial cuff inflation pressure to be inflated initially, its range is different depending on patient
type.
for neonates: initial inflation pressure: setting range: 60.~80 mmHg; default value: 70 mmHg.
for pediatrics: initial inflation pressure: setting range: 80~140 mmHg; default value: 120 mmHg.
for adults: initial inflation pressure:setting range: 80~200 mmHg; default value: 150 mmHg.
Note: if the device is configured with SunTech blood pressure module, then the initial inflation pressure setting
range for adult is 120~200mmHg.
Note: In order to avoid inappropriate initial inflation pressure values which may cause harm to patients, when
the patient type, measuring mode or patient ID is changed, the inflating pressure value will rollback to the
latest setting value.
² "Unit": the pressure unit. mmHg and kPa for optional.
² "NIBP Mode": NIBP measuring mode, "STAT", "Manual", "CUSTOM", "AUTO 1", "AUTO 2", "AUTO 3",
"AUTO 4", "AUTO 5", "AUTO 10", "AUTO 15", "AUTO 20", "AUTO 30", "AUTO 40", "AUTO 50",
"AUTO 60", "AUTO 90", "AUTO 120", "AUTO 240 and"AUTO 480" are optional. When "STAT" is selected,
the device will take a short-term (5 minutes) automatic NIBP measurement. "AUTO 1" means NIBP
measurement takes once every minute automatically; "AUTO 480" means NIBP measurement takes once
every 480 minutes automatically; In AUTO mode, the count-down timer is displayed in MAP(Time) segment
on upper right corner.
Note: when "STAT" (short-term automatic NIBP measurement) mode is selected, the LED display segments
for MAP value will change to display "STAT" confirming the current NIBP mode, therefore the MAP value
will not be displayed. When the "STAT" mode finishes after 5 minutes (or measurement error occurs or it is
interrupted manually), then the device will shift into "Manual" mode automatically.
² "Lock": select "√" means that the cuff inflation pressure is locked at the initial setting value. After "Lock"
item is selected, the cuff inflation pressure will not adapt according to the last NIBP measurement result, it will
be fixed at the preset initial inflation pressure when you make the following NIBP measurement.
² "More >> ": page down icon. Move the cursor to the last item ("Lock"), then short press the Down "
to enter into NIBP verification setup screen, as shown in figure 4.14.