Installation; Opening The Package And Check; Connecting The Power Supply; Starting The Monitor - Gima VITAL PRO-SNET Manual Del Usuario

Ocultar thumbs Ver también para VITAL PRO-SNET:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

3.3 Installation

3.3.1 Opening the Package and Check

1. Open the package, take out the monitor and accessories from the box carefully and place them on a safe sand
table and surface.
2. Open the accompanying document to sort the accessories according to the packing list.
u Inspect the monitor for any mechanical damages
u Check all the accessories for any scratch or deformity, especially on connectors, wire and probe parts
☞ You can customize the module configuration by choosing necessary modules to meet your own needs.
Therefore, your monitor may not have all the monitoring functions and accessories.
If in doubt, please contact the local dealer or our company in case of any problems. We are to offer you the best
solution for your satisfaction.

3.3.2 Connecting the Power Supply

1. When powered by AC mains power supply:
u Make sure that the AC power supply is 100-240VAC, 50/60Hz.
u Use the power cable prepared by the manufacturer. Insert one end of it to the power port of the monitor
and the other end to the grounded three-phase power jack.
u To eliminate potential differences, the monitor has a separate connection to the equipotential grounding
system. Connect one end of the provided ground cable to equipotential grounding port on the rear of the
monitor, and connect the other end to one point of the equipotential grounding system.
Caution: ensure that the monitor is grounded correctly.
% If the mains power is interrupted and restored after 30 seconds (with the mains power switch still on) the
monitor will retain the last settings when the monitor restarts.
2. When powered by built-in battery
uCaution: Recharge the battery while it is almost exhausted wherever possible, the charging time should be
13~15 hours long.
uThe built-in battery of the monitor must be recharged after transportation or storage. So if the monitor is
switched on without being connected to the AC power supply, it may not work properly due to
insufficient battery capacity.

3.3.3 Starting the Monitor

When the monitor is switched on, the system performs a self-test and then enters the initial display. The orange alarm
indicator blinks to inform that the user can begin operating it.
u Check all the applicable functions to make sure that the monitor works normally.
u If the battery is used please recharge it after using the monitor to ensure sufficient power storage. It will
take minimal 8 hours to charge battery from depletion to 90% charge.
% Do not use the device to monitor the patient if there are indications of damage or error message. In this event,
please contact the local dealer or our company.
% Wait one minute before restarting the monitor after turning it off.

3.4 Sensor Placement and Connection

3.4.1 ECG Cable Connection

ECG measurement is to collect the ECG signal via the ECG electrodes. Electrode connects the patient and the lead. The
lead connects the monitor. The locations of the electrodes are very important for obtaining accurate ECG signals.
1. Connect the cable to the right-panel connector marked with the ECG icon.
9

Publicidad

Tabla de contenido
loading

Tabla de contenido