Alarm Reset And Silence; Alarm Settings - Gima VITAL PRO-SNET Manual Del Usuario

Ocultar thumbs Ver también para VITAL PRO-SNET:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

5.3 Alarm Reset and Silence

Press
(Alarm Silence) key to pause the audible alarm temporarily or reset the current alarm condition. During the
monitoring process, press "Alarm Silence" key shortly to start the alarm silence for 2 minutes. The counting down time
shows up on the upper left corner of the screen once the alarm silence is activated. Long pressing "Alarm Silence" key
will reset the current alarm condition, that means the sound mute for this alarm will not resume unless another alarm
condition occurs. During the alarm silence period, if there is a new alarm condition other than the current one occurs, the
device will generate the audible alarm indication again automatically. After the end of alarm silence period, if the current
alarm condition still exists, then the audible alarm indication will resume as well.
When the monitor generates alarms, the user can press
given silence period when necessary..
M DO NOT silence the audible alarm or decrease its volume if patient safety could be
compromised.
M For the alarm conditions of "Can not detect SpO
alarm indication will only last for about 7 seconds .
M Alarm signal can be reset, but it can NOT be deactivated all the time.

5.4 Alarm Settings

1、 Except volume of audible alarm can be adjustable, the other properties of the alarm signal cannot be adjusted
by the user, such as alarm priority setting, alarm light flashing and so on. In addition, all alarms in this patient
monitor are "non-latched" type, that is to say, when the alarm condition does not exist, the corresponding
alarm signal will automatically stop.
The alarm volume range is shown as below:
²
High: 45dB~80dB (The distance from the front of device to the test instrument is 1m)
Medium: 45dB~75dB (The distance from the front of device to the test instrument is 1m)
²
²
Low: 45dB~70dB (The distance from the front of device to the test instrument is 1m)
2、 When the icon
mute), at this time the user should pay more attention to the patient.
It is suggested that the users should not change the alarm volume lower than the factory default setting if close
%
and constant attention could not be paid to the patient, otherwise the negligence of alarm event might cause
irreversible harm to the patient.
During the alarm silence period, any new alarm event can activate the audible alarm again and the audible
%
alarm function resumes normal state.
%
After the alarm silence time counts down to 0, or the operator presses the Alarm Silence Key again, then the
system will resume to the audible alarm signal if this alarm condition still exists.
%
The alarm limit value should NOT be set to exceed the declared measuring or display range, or the system
alarm signal will not be generated.
3、 Alarm settings are non-volatile, that means the previous settings will still sustain if the patient monitor is
powered off (by accidental power interrupt or by normal power down) and reboot.
4、 When pressing the Alarm Silence Key, the system will stay on "Alarm Silence" status and this status will last
for 2 minutes.
5、 It takes only 1 second from canceling alarm silence to resuming alarm sound .
Long press of Display View "
displays on the screen and its color is red, that means the alarm volume is 0 (alarm is
" key to enter into the Setup Menu screen, and move the cursor to each parameter
key to reset or pause the audible alarm indication for a
" and"Can not detect HR/PR", the audible
2
43

Publicidad

Tabla de contenido
loading

Tabla de contenido