4.9.10 System Setup

Screen Description:
² "Alarm Vol.": sets alarm volume, adjustable from leve"1~10" with a factory default of 5. It is recommended
that the alarm volume shouldn't be set lower than the factory default value unless the nursing personnel attends
the patient and the device at all times.
² "Key tone": turns on/off the key tone, the default is "ON".
² "Language": language selection. "ENG" for English.
² "Priority": this item is nonadjustable, fixed to give priority to the "HR" value display.
² "Run mode": "Real" should be set as default use. "Demo" only for demonstration purpose. Changing this item
requires a password, the default password is "1234".
"Demo" shows a demonstration waveform and data, which are generated by the monitor.
"Real" shows the real signal waveform and data coming from patient, i.e. normal working state.
² "Beat beep": adjusts the volume of the pulse beeping sound and is adjustable from level "0~7". "0" switches
off the pulse beep sound, the factory default is set at"2". The tone of pulse beat beep changes when the
measured SpO
2
sharper); and the lower the SpO
Figure 4.22A System Setup Screen
Figure 4.22B System Setup Screen
changes i.e. the higher the SpO
value is, the lower the tone of the pulse beep.
2
value is, the higher the tone of the pulse beep (which becomes
2
38