Dear Users,
Thank you very much for purchasing our product. Please read the following information very
carefully before using this device.
Read these instructions carefully before using this monitor. These instructions describe the
operating procedures to be followed strictly. Failure to follow these instructions can cause
monitoring abnormity, equipment damage and personal injury. The manufacturer is NOT
responsible for the safety, reliability and performance issues and any monitoring abnormality,
personal injury and equipment damage due to user's negligence of the operation instructions. The
manufacturer's warranty service does not cover such faults.
M WARNING-PACEMAKER PATIENTS. This monitor may continue to count the
pacemaker rate during occurrences of cardiac arrest or some arrhythmias. Do not rely
entirely upon
surveillance.
M Monitoring a single person at a time.
M The monitor is defibrillator proof. Verify that the accessories can function safely and
normally and the monitor is grounded properly before conducting defibrillation.
M Disconnect the monitor and sensors before MRI scanning. Use during MRI could cause
burns or adversely affect the MRI image or the monitor's accuracy.
M If you have any doubt to the grounding layout and its performance, you must use the
built-in battery to power the monitor.
M All combinations of equipment must be in compliance with standard of IEC 60601-1-1
for medical electric system requirements.
M Check SpO
2
circulation, positioning and skin sensitivity.
M The SpO
probe of this monitor may not work for all testees. If stable readings can not
2
be obtained at any time, change appropriate probe or discontinue use of SpO
monitoring.
M Do not immerse the monitor or its accessories in liquid to clean.
M Do not use accessories other than those provided/recommended by the manufacturer.
M Each time the monitor is used, check the alarm limits to ensure that they are
appropriate for the patient being monitored.
M The monitor is intended only as an adjunct in patient assessment. It must be used in
conjunction with clinical signs and symptoms.
M When taking the measure of an pediatric or neonate's (less than 10 years old) blood pressure, do
NOT operate in the adult mode. The high inflation pressure may cause lesion or even body
putrescence.
M The monitor is prohibited from applying to those who have severe hemorrhagic tendency or who are
with sickle cell disease for they may develop partial bleeding when this monitor is used to take the
blood pressure measurement.
Instructions to User
this monitor ALARMS. Keep pacemaker patients under close
probe application site periodically (every 30 minutes) to determine
II