Use the clip type test lead with integral temperature sensor to display the battery
terminal on the Tester.
6.
Read the battery internal resistance, DC voltage and temperature directly from the
display.
7.
Press "
Note
When the measured DC voltage or battery internal resistance value is
overloaded, "OL" is displayed
If the resistance reading shows "- - - -" an AC test current fault has occurred
and a measurement could not be made. This may be due to a break in the
test circuit, or due to the leads not making a good contact with the object
being measured. This may also happen if the resistance is extremely large
compared to the measurement range.
4-2 Zero Adjust Function
The zero adjustment function is used to zero the resistance range. The reading taken
during the zero adjust procedure will be taken as zero and will be used to calibrate
subsequent measurements.
4-2-1 Probe -type Test Lead Zero Adjustment
Using the supplied zero adjustment board, zero adjustment can be carried out
using the AC four terminal method.
09/15/17 Version No. 00
" key for 2 seconds to change the displayed temperature unit.
Battery Capacity Tester / EN
10