6. BATTERY CHARGE/DISCHARGE CAPACITY MEASUREMENT
1. Disable auto power off function (see section 4-6).
2. Perform the DC current measurement (see section 4-5 in steps 1 – 3).
3. Connect the fully discharged battery to the battery charger. Clamp the DCA
current adaptor transducer jaws around the "+" terminal.
4. Power on the battery charger, the meter will display the charge current value with a
plus sign and the battery accumulated charging capacity (IN–CAP).
5. Connect the battery load to the battery. Clamp the DCA current adaptor
transducer jaws around the "+" conductor.
6. The meter will display the discharge current value with a minus sign and the
battery accumulated discharge capacity (OUT–CAP).
7. The battery charge / discharge remaining capacity is :
(T–CAP) = (IN–CAP) – (OUT–CAP)
09/15/17 Version No. 00
Battery Capacity Tester / EN