Battery Capacity Tester / EN
1.
Turn the resistance range switch to the desired position.
2.
Press "Zero" key to start the zero adjustment function, the flashing " 0ADJ "
symbol is displayed.
3.
Push the pin type test leads onto two holes in the zero adjustment board (see
illustration below). Choose holes symmetrically on both sides of the centre screw,
so that the distance between the leads is about equal to the distance when
measuring the battery terminals.
4.
When the resistance reading is lower that 200 digits and stable, the zero
adjustment is automatically carried out, the " 0ADJ " stops flashing and the
beeper sounds.
5.
Press "Zero" key to exit this function.
6.
Remove the pin type test leads from the zero adjust board and connect the leads
to the battery to start taking measurements.
Note
If the terminal spacing of the battery to be measured is larger than the zero
adjustment board, use the outer most holes for adjustment.
The zero adjustment is valid for the currently selected range for as long as the
power remains on. Turning the meter off will reset all zero adjustment values.
4-2-2 Clip – type Test Lead Zero Adjustment
1. Short the SOURCE and SENSE of the clip type test lead with temperature sensor
together as shown in the figure below.
2. The rest of the procedure is the same as the probe-type test lead zero adjustment.
11
About equal to the distance when
measuring the battery terminals
Contact with SENSE
Contact with SOURCE
09/15/17 Version No. 00