4-3 HOLD and Auto – HOLD Functions
4-3-1 Hold function
1. Press HOLD key, " HOLD " appears on the display and display is locked.
2. Press HOLD key again to exit.
4-3-2 Auto – hold function
1. Press HOLD key for 2 seconds to activate the auto-hold function, " AUTO-HOLD "
appears on the display.
2. The meter holds the reading on the display until it detects a new stable reading.
The meter then beeps and displays the new reading. When the measured
readings are stable, " HOLD " appears on the display.
3. Press HOLD key again to exit.
4-4 Moving Average Function
If the resistance measurement value is unstable, this can be corrected with the
moving average function.
1. Press and hold down TIME key, then rotate the resistance range switch to any
range to turn on the meter, to samples select mode, "SET" appears on the display.
2. Press
of "4", "8" and "16" are allowed. To stop the averaging function select "OFF"
3. Press TIME key again to store the setting and exit.
4-5 DC Current Measurement
1. Connect the yellow miniature plug of the DCA current adaptor to the meter DCA
CLAMP ADAPTOR jack.
2. Press POWER key of the DCA current adaptor to turn on the adaptor, the
POWER LED will come on. If the LOW BATTERY LED comes on, the battery
should be replaced.
3. Make sure the current adaptor jaws are empty, then press ZERO key to null out
stray magnetism, the LCD should then show 0.0A.