8. Configuration via display and operating unit
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Scale setting" and confirm with [↵].
3. Select "Dry adjustm." and confirm with [↵].
4. Define the start of measuring range or end of measuring
range:
To define start of measuring range
Confirm "min. adjustm." with [↵].
To define end of measuring range
Confirm "max. adjustm." with [↵].
5. Change digit using [▲] [▼] and confirm with [↵]. The cursor
moves to the next digit. Repeat for each digit.
When the last digit is exited, the cursor moves to the output
signal (step 6).
6. Change digit using [▲] [▼] and confirm with [↵]. The cursor
moves to the next digit. Repeat for each digit.
When the last digit is exited, the menu moves back to step 2.
With the input of current values that are not either 4 mA or 20 mA the
pressure value is converted into the standardised current signals as
soon as the current value entered is accepted.
8.4 Setting the mode
The mode defines which measurand will be transmitted via the current output (pressure,
volume).
Level and volume modes cannot be used for this instrument.
If the main display is set to PV (primary value), the measurand set under
"Mode" will always be displayed.
WIKA operating instructions, model DMSU22SA
EN
31