13. Faults / 14. Dismounting ...
Error code
Error text
1)
E006
Sensor overpres-
sure
E007
Sensor tempera-
ture
E008
Diaphragm rupture
1) An error message can also appear if the pressure is greater than the nominal measuring range.
If faults cannot be eliminated by means of the measures listed above, shut
down the instrument immediately, and ensure that pressure and/or signal are
no longer present, and secure the instrument from being put back into opera-
tion inadvertently.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in
chapter 14.2 "Return".
14. Dismounting, return and disposal
WARNING!
Physical injury
Residual media in the dismounted instrument can result in a risk to persons,
the environment and equipment.
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
Wear the requisite protective equipment, see chapter 3.5 "Personal protec-
tive equipment".
WARNING!
Risk of burns
During dismounting there is a risk of dangerously hot media escaping.
Wear the requisite protective equipment, see chapter 3.5 "Personal protec-
tive equipment".
Let the instrument cool down to room temperature before dismounting it.
WIKA operating instructions, model DMSU22SA
Causes
Pressure sensor overload
Temperature exceeded, limit
monitoring of the tempera-
ture at the sensor tube
Sensor break detected,
leakage at the sensor tube
Measures
Restart the instrument
Depressurise the instru-
ment (ambient pressure)
and restart
Return the instrument
Return the instrument
Return the instrument