8. Configuration via display and operating unit
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Application" and confirm with [↵].
3. Select "Mode" and confirm with [↵].
EN
4. Select measurand and confirm with [↵].
» Mode is set.
8.5 Mounting correction (offset)
8.5.1 Performing a wet adjustment
Zero point will be taken from the running measurement.
Requirement:
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Application" and confirm with [↵].
3. Select "Pressure" and confirm with [↵].
4. Select "Mounting corr." and confirm with [↵].
5. Select "apply" and confirm with [↵].
Current measured value will be used as the new zero point.
32
Deviation ≤ 20 % of the measuring range.
Absolute vacuum with absolute pressure measuring instruments.
Not to be carried out without suitable equipment.
WIKA operating instructions, model DMSU22SA