10. Detailed settings
Setting range: 3.8 ... 20.5 mA or 4.0 ... 20.0 mA
(NAMUR recommendation NE43 for process instruments is 3.8 ... 20.5 mA)
1. Open the operating menu with [↵].
Select "Detail setup" and confirm with [↵].
2. Select "Current out" and confirm with [↵].
EN
3. Select "Limits" and confirm with [↵].
4. Select signal limits and confirm with [↵].
» Signal limits are set.
10.5 Setting the contrast of the LC display
Setting range: 1 ... 9 (in steps of 1)
1. Open the operating menu with [↵].
Select "Detail setup" and confirm with [↵].
2. Select "Contrast" and confirm with [↵].
3. Change figure using [▲] [▼] and confirm with [↵].
» Contrast is set.
10.6 Restoring factory setting
1. Open the operating menu with [↵].
Select "Detail setup" and confirm with [↵].
2. Select "Reset" and confirm with [↵].
40
WIKA operating instructions, model DMSU22SA