Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which the RF disturbances are controlled. The customer or user of the device
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
[W]
0.01
0.1
1
10
100
For transmitters whose maximum rated power is not listed in the table above, the recommended separation distance d in meters (m) can be estimated using
the equation from the respective column, whereby P is the maximum rated power of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic waves is affected by absorptions and reflections of buildings,
objects, and people.
Compatibility with surgical HF instruments
This device was tested for compatibility with HF
surgical devices. It has been tested according to
IEC 6060122 Annex BB.
Electromagnetic
Compatibility (EMC)
information
Table 5
Separation distance d [m] according to frequency of transmitter