8.4.1 Changing the prosthesis setting using the cockpit app

Product
MyModes
Functions
Settings
Status
General
Manage components
Imprint/Info

8.4.2 Overview of adjustment parameters in basic mode

The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait
cycle. These parameters act as basic settings for automatically adjusting the damping behaviour
to the current motion situation (e.g. ramps, slow walking speed, etc.).
The following parameters can be modified:
Remote control
parameter
Pitch
Volume
Heel Resist.
Rollover Resist.
Stair Function
1) Once the component is connected and in the desired mode,
tap the  symbol in the main menu.
→ The navigation menu opens.
change
2) Tap the menu entry "Settings".
→ A list appears with the parameters for the currently
2. Golf
3) Change the setting of the desired parameter by tapping the
"<", ">" symbols.
INFORMATION: The prosthetist's setting is marked
and, after the setting has been changed, can be
restored by tapping the "Standard" button.
Adjustment
software
range
1000 Hz —
4000 Hz
0 — 4
10 — 60
110 — 170
ON — OFF
selected mode.
App/remote
Meaning
control
adjustment
range
1000 Hz —
Pitch (frequency) of beep signal for
4000 Hz
acknowledgement tones
0 — 4
Volume of beep signal for confirma­
tion tones (e.g. when checking the
charge level, switching MyModes).
The "0" setting deactivates the aud­
ible feedback signals. However, warn­
ing signals in case of errors are still
generated (see Page 68).
± 20
Plantar flexion damping.
How fast the forefoot is lowered when
weight is placed on the heel.
± 10
This parameter defines how easy the
rollover is.
ON — OFF
Switching on this function increases
the rollover angle when walking down
stairs. This function needs to be
enabled in the adjustment software.