Chlorine Probe Calibration Menu; Redox Probe Calibration Menu - Aqua SALT Cella 50 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

AQUA SALT
Salt chlorine generator
The Menu "Calibration of chlorine probe" allows you to perform the calibration of the chlorine probe, if the
4.5.3
Aqua Salt model provides for their presence. Calibration should be performed before using the sensor
Chlorine
using the following procedure:
probe
1)
The probe must be inserted by at least two hours in the probe holders under the following conditions:
calibration
Flow of 30 l / h, 0.8-4 ppm chlorine level, pH range 4-12 pH, pressure less than 0.5 bar. For proper
installation of the probe see the Addendum to this manual, specifically for the CL-J probe.
menu
2)
From the screen that prompts you to insert the probe into the water, press ENTER.
3)
You will see a countdown at the end of the countdown or after you press the Enter key shows the
chlorine value read by the probe expressed in mA. You can leave out this information. A probe
functioning correctly (reading range 0-5 ppm) roughly follows the following equation: mA = 4 mA +
3.2mA / ppm.
4)
after 5 seconds, it displays the value of the measured chlorine (uncalibrated) from the probe and
expressed in ppm.
5)
If you press the Enter key returns to the value of the previously measured chlorine. Use the Up and
Down keys to change this value to the value measured with the photometer (DPD method).
Pressing for about a second the Enter key to return to the installer menu item Calibration Chlorine.
The Redox Probe Calibration Menu calibrates the Redox probe in case the model supports it.
4.5.4
calibration regulates the measurement tool and improves its efficiency. It must be performed before using the
Redox
probe, following the step-by-step procedure below:
probe
1) Rinse the probe and dip it in the buffer solution at 475mV, move slowly for 10 seconds and press
calibration
Enter.
2) Wait without touching the probe or its cable for at least 60 seconds, until the value read is
menu
stabilized.
3) At the end of the countdown or after pressing Enter, the Redox value of the buffer solution (475
mV) and the value read by the probe will be shown.
4) The procedure ends visualizing the quality of the probe with a percentage from 0 to 100. If quality
is below 25%, we recommend to replace the probe.
5) Place the probe in the probe holder or in the filtration system.
Press Enter to scroll the menu sections.
Press and hold Enter for about 1 second to go back to the Redox Calibration section of the Installer Menu.
ADSP7000648
Dip the probe
in the water
Set chlorine
read chlorine
Dip the probe in
the water
Probe Quality
Rx 475 mV
rev. 1.4
MAINTENANCE AND INSTRUCTION MANUAL
Waiting
Chlorine mA
Waiting
31/07/2017
ENGLISH
78
Probe
78/264