7.1.3

Version II: Passive / passive current outputs

Model FCx3xx-A1, FCT3xx-A1 or FCx3xx-A2, FCT3xx-A2 or FCx3xx-S2, FCT3xx-S2
Current output 1, passive
Terminals 31 / 32
Current output 2, passive
Terminals 33 / 34
Pulse output, passive
Terminals 51 / 52
Switching output, passive
Terminals 41 / 42
Switching input, passive
Terminals 81 / 82
All inputs and outputs are electrically isolated from one other and from the auxiliary power.
Change from one to tw o c olumns
7.1.4

Special connection conditions

The output circuits are designed so that they can be
connected to both intrinsically-safe and non-intrinsically-safe
circuits.
It is not permitted to combine intrinsically safe and non-
intrinsically safe circuits. In case of a change of type of
protection observe chapter 6.6.7.
On intrinsically-safe circuits, equipotential bonding must be in
place along the entire length of the cable used for the current
outputs.
The rated voltage of the non-intrinsically safe circuits is
U
= 60 V.
M
The switching output and the pulse output (terminals 41 / 42
and 51 / 52) can be wired internally as a NAMUR contact for
the purpose of connecting a NAMUR amplifier.
Change from two to one c olumn
Type of protection
General operating
"nA" (Zone 2)
values
U
(V)
I
(mA)
U
(V)
i
i
b
30
30
30
30
30
30
30
65
30
30
65
30
30
10
30
Type of protection
"e" (Zone 1)
I
(mA)
U (V)
I (mA)
b
30
60
35
30
60
35
65
60
35
65
60
35
10
60
35
The cable glands are supplied in black by default. If the signal
outputs are wired to intrinsically-safe circuits, we recommend
that you use the light blue caps supplied for the appropriate
cable entries.
IMPORTANT (NOTE)
If the protective conductor (PE) is connected in the
flowmeter's terminal box, you must ensure that no dangerous
potential difference can arise between the protective
conductor (PE) and the equipotential bonding (PA) in the
hazardous area.
CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-X1 Rev. F EN - 35
Type of protection "ia"
(Zone 1)
U
I
P
C
i
i
i
i
(V)
(mA)
(mW)
(nF)
60
300
2000
0.47
60
300
2000
0.47
60
300
2000
0.47
60
300
2000
0.47
60
300
2000
0.47
C
pa
L
i
i
(nF)
(mH)
0.47
0.17
0.47
0.17
0.47
0.17
0.47
0.17
0.47
0.17