1463021000_00:R12895/70080272
FEEDING ADMINISTRATION
WARNING: Do not fill balloon with medications.
1. Remove the feed port plug (Figure 3,A).
2. If feeding with a syringe, connect the catheter tip syringe to the bolus extension feed port (Figure 4,A). Push firmly and twist
1/4 turn to secure the connection.
3. If using a feeding bag, purge the air from the bag and tubing. Connect the tubing set to the SECUR-LOK Extension Set
(Figure 3,B). Push firmly and twist 1/4 turn to secure the connection.
4. Adjust the formula flow rate and administer the feeding.
5. Upon completion, flush the extension set and MIC-KEY with 20 ml warm water until the tubing is clear.
6. Disconnect the SECUR-LOK or the Bolus Extension Set and replace the MIC-KEY feed port plug (Figure 2,C).
7. Wash the extension set, and the catheter-tip syringe or feeding bag with warm soapy water and rinse thoroughly.
8. If feeding continuously with a pump, flush the MIC-KEY with 10 ml to 20 ml water every six hours.
DECOMPRESSION
1. Decompression may be done with either the Bolus or the SECUR-LOK Extension Set.
2. Attach the Bolus (Figure 4) or the SECUR-LOK Extension Set (Figure 3) to the MIC-KEY (Figure 2).
3. If using the SECUR-LOK Extension Set, remove the feed port plug (Figure 3,C). Drain the stomach contents into a container.
4. After decompression, flush the extension set and the MIC-KEY with 20 ml warm water.
5. Remove the extension set and replace the attached MIC-KEY plug. (Figure 2,C).
DAILY MAINTENANCE
1. Inspect the stoma site for redness, swelling, discomfort, or gastric leakage.
2. Rotate the MIC-KEY and gently clean the skin around the stoma.
3. Flush the MIC-KEY after each use with 10 ml to 20 ml water.
4. Verify balloon volume once a week. Correct balloon volume is usually 5 ml (3 ml for 12 Fr.). Aspirate all of the liquid from
balloon and refill to the original prescribed amount.
BALLOON LONGEVITY
Precise balloon life cannot be predicted. Silicone balloons generally last 1–8 months, but the life span of the balloon varies
according to several factors. These factors may include medications, volume of water used to inflate the balloon, gastric pH,
and tube care.
REPLACEMENT PROCEDURE
1. Remove water from the balloon with a luer slip Syringe (Figure 2,A).
2. Rotate and gently remove the MIC-KEY from the stoma. Water soluble lubricant may help to remove it.
3. Replace the MIC-KEY following the steps described under Placement Procedure.
MIC-KEY*, MIC* and SECUR-LOK* are trademarks of Kimberly-Clark or its affiliates.
Single Use Only
WARNING: FOR ENTERAL NUTRITION AND/OR MEDICATION ONLY.
KIT CONTENTS:
1 MIC-KEY* Low Profile Feeding Tube
1 MIC-KEY* Extension Set with SECUR-LOK Right Angle Connector and 2 Port "Y" and Clamp-12"
1 MIC-KEY* Bolus Extension Set with Cath Tip, SECUR-LOK Straight Connector and Clamp-12"
1 6 ml Luer-Slip Syringe
1 35 ml Catheter Tip Syringe
4 Gauze Pads
* Registered Trademark or Trademark of Kimberly-Clark Worldwide, Inc.
© 2003 KCWW. All rights reserved.
Single Use
LOT
Lot Number
2
2/21/08 • 14-63-021-0-00/70080272 DFU 0120-12 SERIES – PAGE 2 • 5.5"W x 8.5"L • BLACK
2/21/08
!
STERILE EO
Sterilized Using
Read
Ethylene Oxide
Instructions
REF
Reorder
Number
10:17 AM
Page 4
Do Not Use
If Package
Damaged
Quantity
Rx Only
(USA)
DEHP (Di(2-ethylhexyl) Phthalate)
Length
Free Formulation