Function Test
en
44 / 412
6.3

Scale Test

1.
Turn the system on.
2.
Once the message Insert Tube Set appears, press the Pause/Resume button
and the Zero button simultaneously.
The pump display depicts the message Scale Test.
3.
The first option is the test of the container scale.
4.
Place the 1000 g weight (included with each pump) on the container scale.
5.
The display of the fluid deficit limit depicts the weight. The displayed value
should be 1000 g. The acceptable tolerance is ±20 g.
6.
If the displayed value is outside the tolerance range, the scale must be calib-
rated by a service technician.
7.
Remove the weight from container scale.
Press the Pause/Resume button to conclude this test.
8.
Record results in the test log in the Appendix [}  62]. The test is successful if res-
ults are within the acceptable tolerance limits.