5.
Guide the power cord through the holes provided for this purpose and connect
to pump (1) (Fig. Scale and pump connection [} 27]) and plug into grounded,
shockproof safety wall socket.
6.
Attach the scale to the pump by connecting and screwing the connector of the
container scale (3) (Fig. Scale and pump connection [} 27]) and fix the con-
nected cables below the lower pump tray by means of the provided wire clips.
WARNING!
Scale error
Ensure that nothing weighs down the scale during system start-up. Doing so may
result in an inaccurate deficit value.
WARNING!
Scale error
Make sure that screw connections at cable and pump are fixed hand-tight to avoid
loosening of connection during move or use.
WARNING!
Fluid deficit
The fluid left in the patient must be monitored. The deficit is the total amount of
fluid left in the patient or unaccounted for otherwise. Take notice of the measure-
ment tolerance of the system (see Chapter Technical Data [} 52]). Estimating the
fluid volume remaining in the patient is the physician's responsibility.
Fig. 5–4 Scale and pump connection
(1) Power cord/pump connection
(2) Power output
(3) Plug connection pump with
container scale
(4) Handwheel
(1)
(5) Screw
(2)
(3)
(4)
(5)
System Operation
en
27 / 412