System Operation
en
Precise balancing
Scale capacity
28 / 412
WARNING!
Serum sodium concentration
It is also necessary to monitor the concentration of sodium in the blood of the pa-
tient to prevent electrolyte disturbances. Monitoring of the concentration of so-
dium in the blood must be performed by the physician and is not performed or sup-
ported by the system.
WARNING!
Incorrect measurement of the deficit (inflow tube)
Air located in the inflow tube results in an incorrect measurement of the deficit.
Make sure to replace the bag in a timely manner.
WARNING!
Incorrect measurement of the deficit (bag clamp)
Make sure the clamp of the active bag is always open during operation of the
pump. Otherwise, this will lead to an incorrect measurement of the deficit.
NOTE!
Deficit accuracy
The greater the consumption of irrigation fluid, the greater the deviation between
the actual and displayed deficit (see "Technical Data", accuracy: deficit ±10 %).
Try to collect all the fluid running out of the uterine cavity during the procedure to
achieve the most exact deficit value possible.
The scale can be loaded with a weight of up to 65 lbs (~30 kg). Weight above this
value triggers the Scale Overloaded/Check Scale message. Three audible warning
signals are emitted (see Chapter Error and Warning Messages [} 54]).
WARNING!
Containers and fluid bags
Make sure the containers and fluid bags hang freely, are not resting on something,
and do not touch other objects. Failure to follow these instructions means the defi-
cit cannot be calculated correctly.
NOTE!
Scale recognition
Connect the scale to the pump before turning the system on to ensure the system
recognizes the scale.