Measurement range
Pressure [mmHg]
Deficit [ml]
Inflow volume [ml]
Dimensions
Width x Height x Depth [in]/[mm]
Weight [lbs], [kg]
Accuracy
Flow [% measured value]
Pressure [mmHg]
Deficit [% measured value]
Suction pressure
Interfaces
Signal IN/OUT
Mains power socket
RFID transponder technology
Pressure: build up, measure and control in a body cavity
NC: Maximum value: 150 mmHg ± 10 mmHg
SFC: Maximum value: 200 mmHg ± 10 mmHg for maximum 5 s
Suction pressure High:
Essential Perform-
NC: -67 kPa ± 20 %
ance
SFC: 0 kPa (no function) or -84 kPa ± 20 % (constructional limit)
Suction pressure Low:
NC: -30 kPa ± 20 %
SFC: 0 kPa (no function) or -84 kPa ± 20 % (constructional limit)
1
A tube set is not an application part in terms of the standard. However, it meets
all the technical requirements for an application part.
0 - 500 mmHg
-999/+9999 ml
0 - 30000 ml
12 in x 6 in x 12 in / 300 mm x 140 mm x 300 mm (AQL-100P),
26 in x 52 in x 26 in / 670 mm x 1320 mm x 670 mm
(AQL-100CS)
13 lbs [5.8 kg] (AQL-100P), 23 lbs [10.5 kg] (AQL-100CS)
± 10 %
± 10 mmHg
< 1 l: ±100 ml
> 1 l: ± 10 %
± 20 %
1 x scale connection (round flanged connection socket with 5
pins)
1 x service port (RS232 socket D-SUB9)
IEC-60320-1 C14
Transmit/receive frequency range: 13.56 MHz ± 0.424 MHz
Transceiver class: class I
RF output power: -10.83 dBμA/m at 10 m/32.8 ft
Type of antenna: inductive loop antenna
Antenna loop area: 0.00032 m²
Modulation: amplitude-shift keying (ASK)
Mode of operation (simplex/duplex): duplex