Ivoclar Vivadent Centric Tray Instrucciones De Uso página 3

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 10
Technical instructions
1. Trim back the vestibular edges of the silicone impression. Try to identify three reference points to the models.
2. Seat the models in the silicone impression and check the fit.
3. Mount the models in the articulator, positioning them in relation to the occlusal record.
4. If a UTS Transferbow has been used, attach it to the articulator, applying the philosophy relative to that determined by the Transferbow
record.
5. Secure the Ivoclar Vivadent Centric Tray in place. Place and fix the maxillary model. Remove lower jig and locate and fix the mandibular
model.
6. Fabricate individual trays, centre pin recording devices, wax occlusal rims or other prescription requests.
Please observe the following:
– Keep a record of the measured occlusal height.
– Use a viscous silicone, alginate or similar material.
– Use sufficient impression material.
– Check the seating of the models.
– Consider the articulator philosophy.
Reprocessing
– Autoclavable tray for multiple use (recommended for 10 autoclave cycles. If the tray is subjected to further autoclave cycles, the inherent
stability of the tray may be compromised and thus the exact dimensions are no longer ensured).
– Manual disinfection:
Disinfection procedures:
Disassemble the tray before disinfecting it. Place the Centric Tray into a mesh basket. Immerse the mesh basket with the individual parts of
the tray in disinfectant solution, e.g. MD 520 (Dürr Dental), at 20 ± 2 °C/ 68 ±3.6 °F for 5 min. Make sure that the surfaces of the components
are completely covered with disinfectant solution and that they are free of bubbles during the reaction time. Rinse Centric Tray with cold
demineralized water to remove any disinfectant residues.
– Sterilization: For steam sterilization, use the following parameters: 3x pre-vacuum procedure , 134 °C/273.2 °F, sterilization time 5.5 min.
Safety information
− In the case of serious incidents related to the product, please contact Ivoclar Vivadent AG, Bendererstrasse 2, 9494 Schaan/Liechtenstein,
website: www.ivoclarvivadent.com, and your responsible competent authority.
– The current Instructions for Use are available in the download section of the Ivoclar Vivadent AG website (www.ivoclarvivadent.com).
Disposal information
– Remaining stocks must be disposed of according to the corresponding national legal requirements.
Shelf life and storage
This product does not require any special storage conditions.
Additional information
Keep material out of the reach of children!
The material has been developed solely for use in dentistry. Processing should be carried out strictly according to the Instructions for Use. Liability
cannot be accepted for damages resulting from failure to observe the Instructions or the stipulated area of use. The user is responsible for testing
the materials for their suitability and use for any purpose not explicitly stated in the Instructions.
3

Publicidad

Tabla de contenido
loading

Tabla de contenido