Philips SRO 2550 Manual De Instrucciones página 26

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Conditioning procedure
Switch ON the generator and select an appropriately programmed auxiliary for the
X-ray tube housing assembly to be conditioned.
Select the large focal spot of the X-ray tube housing assembly.
Select 2-button-technique "kV-mAs".
Choose appropriate values at the generator desk corresponding to the following table.
Release exposures according to the following table.
Note:
In case of electrical interferences, the procedure must be continued after an interval of
5 min, recommencing at a lower kV level.
Number of exposures
Table 2: Values for conditioning of the X-ray tube housing assembly
*) Given values depend on the generator.
They may be slightly different.
20
X-ray tube voltage
[kV] *
5
3
102
2
117
2
141
(05.0)
2004 Philips Medical Systems
ALL RIGHTS RESERVED
mAs product
[mAs]
81
125
125
125
125
SRO 2550 ROT 350
Break time
[s]
1
30
1
30
1
60
1
120
SRO 2550 ROT 350
SRO_2550_ROT_350_Manual_English.fm

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Rot 350

Tabla de contenido