F F i i r r e e b b a a r r r r i i e e r r r r e e p p l l a a c c e e m m e e n n t t
T T o o o o l l s s R R e e q q u u i i r r e e d d : :
None
P P r r o o c c e e d d u u r r e e : :
1. Raise the bed height to the full up position.
2. Lower the Fowler and gatch sections to the full down positions.
3. Unzip the cover. The zipper starts in the middle of the patient right side of the support surface.
4. Remove and save the top cover.
5. Starting at the foot end, roll up the fire barrier on the foam crib assembly.
N N o o t t e e - - Work from side to side, a little at a time, to move the fire barrier to the top of the support surface.
6. Discard the fire barrier.
7. Starting at the head end, roll the new fire barrier down and slide the fire barrier over the foam crib assembly.
N N o o t t e e - - Align the fire barrier on the foam crib assembly before you slide the fire barrier over the foam crib assembly.
8. Slide the fire barrier down the foam crib assembly, work from side to side, to make sure that the fire barrier is tight on the
foam crib assembly.
9. Align the foam crib assembly on top of the bottom part of the cover.
N N o o t t e e - - Spread the excess fire barrier material equally below the foam crib assembly at the foot end.
10. Fold and align the top cover over the top of the foam crib assembly.
11. Zip the cover to close.
12. Verify proper operation before you return the product to service.
EN
18
2872-009-005 Rev B.1