Aesculap Power Systems
Acculan® 3Ti attachments
9.7 Inspection, maintenance and checks
Note
Aesculap recommends occasional spraying of moving
parts, e.g. triggers, couplings, cover lids, etc. with
Aesculap STERILIT® Power Systems Oil Spray GB600!
Fig. 10
Fig. 11
20
Allow the product to cool down to room
temperature.
Inspect the product after each cleaning and
disinfecting cycle to be sure it is clean, functional
and undamaged.
Mount the STERILIT® Power Systems adapter
supplied with the product on Aesculap STERILIT®
Power Systems Oil Spray and spray into the
cannulation of the small power drill, see Fig. 10,
and through the drill, reamer and Kirschner wire
attachments, see Fig. 11, in each case from the rear
end and for about 1 sec.
After cleaning/disinfecting, inspect all surfaces and
difficult to access spots on the product for visible
dirt or debris.
Check the product for any damage, atypical
running noise, overheating or excessive vibration.
Inspect tools for broken, damaged or blunt edges.
Set aside the product if it is damaged.
9.8 Packaging
Follow the instructions for use of the respective
packaging and storage holders (e.g. instructions for
use TA009721 for the Aesculap Eccos® holder
system).
Insert the product in its proper position in the
Eccos® holder, or put it on a tray in such a way that
the product is protected against damage. Make
certain that all cutting edges are protected.
Package trays appropriately for the sterilization
process (e.g. in Aesculap sterile containers).
Pack the product in such a way that the packaging
will prevent recontamination of the product
(DIN EN ISO 11607).