Mechanical Cleaning/Disinfecting - B.Braun Acculan 3Ti Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 78
Aesculap Power Systems
Acculan® 3Ti attachments
Stage V
Dry the product with lint-free tissue or medical-
quality filtered compressed air.

9.6 Mechanical cleaning/disinfecting

Note
The disinfector must be of proven effectiveness (e.g.
DGHM or FDA approval or CE mark).
Note
For thermal disinfection, always use fully desalinated
(demineralized) water. Ensure that Ao is >3 000 for the
process.
Note
The disinfector used for processing must be serviced
and checked at regular intervals.
If the washer/disinfector is not fitted with internal
rinsing, carry out manual precleaning of the
cannulation, see Manual cleaning/disinfecting,
Stage I.
For Kirschner wire attachment GB643R: Set
adjustment sleeve 4.2 to the largest Kirschner wire
diameter.
For sagittal saw attachment GB660R: Do not rinse
through the interior of the sagittal saw
attachment. Do not connect the Eccos® holder of
the sagittal saw attachment to the rinsing
connector of the washer/disinfector.
Use the Eccos® tray fitted with Acculan® 3Ti
holders, or mount appropriate holders in a suitable
tray (instructions for use TA009721 for Aesculap
Eccos® holder system).
Insert the product in its proper position in the
Eccos® holder.
Connect the internal rinsing equipment at the
Eccos® holder and at the special connector of the
rinsing cart.
18

Publicidad

Tabla de contenido
loading

Tabla de contenido