9 Fault Display and Fault Elimination
9.1 LED flashes red and beeper sounds for 3 seconds
Probable
Error in the Axon-Bus prosthetic system.
cause:
Remedy:
1. Switch the Axon-Bus prosthetic system off and back on.
2. Check the plug connections in the socket.
3. Connect the charger. If the charger displays the service symbol (yellow
9.2 The LED flashes three times after pressing the button in the colour of the current
battery capacity
Probable
The connection from the AxonEnergy Integral to the Axon-Master has been
cause:
interrupted.
Remedy:
1. Check the connection (e.g. plug connections, flex cable, EasyPlug etc.).
2. If no fault can be found, send the Axon-Bus prosthetic system to Ottobock
9.3 LED alternately flashes yellow/red and beeper sounds for 3 seconds
Probable
The Axon-Bus prosthetic system is overheated.
cause:
Remedy:
1. Allow the Axon-Bus prosthetic system to cool for a few minutes.
2. Do not use the Axon-Bus prosthetic system at high ambient temperatures.
9.4 The movements of the Axon-Bus gripping component are becoming increasingly
slower
The patient is alerted to the diminishing battery charge level as the Axon-Bus gripping component
increasingly slows down and the gripping force decreases. Finally, when there is very little battery
capacity remaining, the Axon-Bus prosthetic system switches off automatically to protect the bat
tery against harmful deep discharge.
16 | Ottobock
1x long
spanner) or the system error symbol (red spanner) send the Axon-Bus pros
thetic system to Ottobock Myo-Service.
Myo-Service.
Beep
Beep
2x short
Beep
Beep
3x short
1x short,
1x long
Beep
pulsating