S S e e t t u u p p
Stryker Endoscopy considers instructional training an integral part of the SDC3 system. Your
Stryker Endoscopy sales representative will perform at least one in-service at your convenience to
help you set up your equipment and instruct you and your staff on its operation and maintenance.
Please contact your local Stryker Endoscopy representative to schedule an in-service after your
equipment has arrived.
1. Choose a location for the SDC3
Place the SDC3 on a Stryker cart or other sturdy platform.
Provide a minimum of 4 to 6 inches of space around the sides and top of the SDC3. This
allows for sufficient ventilation and access to the power cord.
W W A A R R N N I I N N G G - - RF and other mobile communications equipment may affect the normal function
of the SDC3. When placing the SDC3, follow the instructions located in the "Electromagnetic
Compatibility" section of this manual.
2. Connect peripheral devices according to the interconnection diagram on the following pages.
W W A A R R N N I I N N G G - -
Connect only devices that have been specified as part of, or compatible with, the SDC3
system.
When the SDC3 is interconnected with other electrical devices, leakage currents may be
additive, resulting in electromagnetic emissions that can interfere with the normal function
of electronic medical equipment. To properly control electromagnetic emissions and avoid
potential harm to the patient or user, ensure all electrical devices are installed and
interconnected according to the requirements of IEC 60601-1 and IEC 60601-1-1.
When installed inside the patient environment, represented by an area 1.5 meters from the
patient in all directions, the printer must be connected to an isolation transformer prior to
use. Recent Stryker carts include isolation transformers and printers can be installed
without the use of an additional transformer. Please contact your Stryker Endoscopy sales
representative to determine if your cart includes this feature.
3. Connect the AC power.
Connect the provided hospital power cord to the AC inlet on the rear console panel.
Connect the other end to a hospital-grade power outlet.
Use a hospital power cord, part number 0105033001 or part number 0105150027.