RF emissions CISPR 11
Harmonic emissions IEC61000-3-2
Voltage Fluctuations/ flicker
emissions IEC61000-3-3
Guidance and Manufacturer's Declaration–Electromagnetic Immunity
The SDC3 system is intended for use in the electromagnetic environment specified below. The user of the SDC3 system should
ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
6 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
SDC3 is intended for use in the electromagnetic environment specified below. The customer or the user of SDC3 should ensure
that it is used in such an environment.
Immunity Test
Electrostatic Discharge (ESD)
IEC61000-4-2
Electrical fast transient/burst
IEC61000-4-4
Surge IEC61000-4-5
P23566L
Class B
Class A
Complies
Compliance Level
6 Vrms
3 V/m
IEC 60601 test level
Compliance Level
±8kV contact
±8kV contact
±15kV air
±15kV air
±2kV for power
±2kV for power supply lines
supply lines
±1kV for input/output lines
±1kV for input/output
lines
(if applicable)
(if applicable)
±1kV differential
±1kV differential mode
mode
±2kV common mode
±2kV common mode
SDC3 is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes, provided the
following warning is heeded:
W W A A R R N N I I N N G G — — This system may cause radio interference or
may disrupt the operation of nearby equipment. It may be
necessary to take mitigating measures such as reorienting
or relocating the system or shielding the location.
Electromagnetic Environment–Guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the SDC3 system, including its
cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance:
d = 2√P
80 MHz to 2.7 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey (a), should be less that the
compliance level in each frequency range (b).
Interference may occur in the vicinity of equipment marked with
the following symbol:
Electromagnetic
Environment - guidance
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial or
hospital environment