W W a a r r n n i i n n g g s s a a n n d d C C a a u u t t i i o o n n s s
Please read this manual and follow its instructions carefully.The words warning, caution, and note
carry special meanings and should be carefully reviewed:
W W a a r r n n i i n n g g : : Indicates measures to avoid potential serious injury to the user and the patient and/
or damage to this device.
C C a a u u t t i i o o n n : : Indicates risks to the equipment. Failure to follow cautions may result in product
damage.
N N o o t t e e : : Provides special information to clarify instructions or present additional useful
information.
1. Caution: Federal law (United States of America) restricts this device to use by, or on order of, a
physician.
2. Read this operating manual thoroughly and be familiar with its contents prior to setting up or
using this equipment.
3. Carefully unpack the unit and check for any damage that may have occurred during shipment.
If damage is detected, contact your Stryker representative.
4. Use of this device is restricted to qualified health care professionals who have a complete
knowledge of the use of this equipment and the devices to which it can be connected.
5. Test this equipment prior to using it in a surgical procedure. This unit was fully tested at the
factory before shipment; however, it must be tested for proper function in the environment and
setup configuration in which it will be used.
6. To avoid risk of electric shock, use only a hospital-grade power cord furnished with the unit.
Disconnect the unit from the power supply to make connections or to inspect the equipment.
7. To avoid the risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
8. Do not position the console so that it is difficult to disconnect the power cord from the supply
mains.
9. Install this device in an operating room that complies with all applicable IEC, CEC, and NEC
requirements for safety of electrical devices. Any installation or connection with other devices
shall be evaluated for electrical safety according to the IEC 60601-1 standard.
10. Use only clean, dry, undamaged Stryker brand media. The use of damaged, wet, or
incompatible media will void the warranty and may result in the malfunctioning of this
equipment.
11. Make sure that all equipment connected to the SDC3 is properly grounded (and isolated, if
required).
12. To avoid risk of fire, do not use this device in the presence of flammable anesthetics.
13. Avoid direct exposure to the laser beam of the DVD/CD drive as it may harm the eyes.
14. Pay close attention to the care and cleaning instructions in this manual. Failure to follow these
instructions may result in product damage.
15. Do not modify this equipment without authorization of the manufacturer.
16. Attempt no internal repairs or adjustments not specifically detailed in this operating manual.
Refer any adjustments, modifications, and/or repairs to Stryker Endoscopy or its authorized
representatives.
17. This equipment complies with FCC radiation exposure limits set forth for an uncontrolled
environment. This equipment should be installed and operated with minimum distance 20 cm
between the radiator and your body.
P23566L