W W A A R R N N I I N N G G - -
Check the power cord assembly periodically for damaged insulation or connectors.
To avoid risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
4. (Optional) Connect the equipotential ground probe to an equipotential bus bar (if available).
The ground probe is provided to make sure all devices installed on the cart are connected to
the same ground potential. Make sure all connections are made according to the requirements
of IEC 60601-1.
5. (Optional) Connect the SDC3 to the WiSe Receiver.
Insert the LED token provided with the WiSe receiver into the token slot on the front of the
SDC3 console according to the sequence of steps described in the WiSe receiver user
guide.
When the LED light is off, no link is active, and the token can be inserted to begin
synchronization.
When the LED light is green, the link is active, and it is okay to remove the token.
When the LED light is amber, the system is busy and the token should not be inserted or
removed.
N N o o t t e e - - The SDC3 can only be linked to one WiSe Receiver.
6. (Optional) Connect the SDC3 to the SYNK Transmitter.
Connect a SIDNE cable to the RS232 port at the back of the SYNK transmitter.
Connect the other end of the SIDNE cable to one of the Devices USB ports at the back of
the SDC3.
From the Current Case > Controls screen, go to the Transmitter tab.
Press the Link button to link the SYNK transmitter and the SDC3. All controls for the
transmitter are now available through the SDC3 touchscreen.
W W A A R R N N I I N N G G - - In the event of power loss, all DVI output signals routed from the SDC3 to the
surgical display will be lost. Analog signals (RGBHV, S-Video, Composite) will not be affected by
power loss, and can therefore be used as backup video signals if they have been connected prior
to power loss.