Interface Lock
The interface lock prevents any unwanted changes to the control unit settings.
Press and hold the alarm mute button
Confirmed by the padlock LED
When the interface is locked, only the alarm mute button is operational.
To deactivate, press and hold the alarm mute button
Deactivation is confirmed by the padlock LED
Alarm Mute Control Button
When activated, the alarm can be paused for up to 10 minutes while troubleshooting, after which time, the
alarm recommences if the problem remains.
To mute the alarm, press the alarm mute button
If a new alarm condition occurs when paused, the initial alarm terminates and a new alarm will activate.
Alarms and Alerts
The Toto® Touch control unit has three integrated alarms and one information alert to ensure patient safety
and efficacy.
Alarms and Alerts
Air Leak Alarm - The pressure in the
system has fallen below the minimum
operational requirements, i.e. an air leak.
Blockage Alarm - Airflow between the
control unit and the platform is prevented,
i.e. a blocked air hose.
Service Interval - Service required; default
at one-year use.
Power Loss Alarm
Power to the control unit has been
interrupted.
Initialisation Failure
Control unit fails to start
The sound pressure level of the alarm audio or the reminder audio signal ranges from 55dB-85dB.
The alarm frequency is 2000HZ.
While troubleshooting, the alarm can be muted for up to 10 minutes. If the fault is not corrected during this
time, the alarm re-activates.
If a new alarm condition occurs within 10 minutes, mute will cease, and the new condition will generate an
alarm.
Please Note:
The corresponding alarm indication LED remains lit when the alarm is muted.
If the alarm condition is corrected while muted, the LED and mute will automatically terminate.
The alarm audio signal comes from a speaker, and the reminder audio signal comes from a buzzer.
Alarm activation may take up to two minutes from when the alarm state occurs.
for four seconds.
and a double beep.
INFLATE TIME
L & R
R
touch
L
Alarm
Condition
Low priority
Low priority
n/a
Low priority
n/a
for four seconds.
turning off, and a double beep.
.
240 min
180 min
30 min
120 min
60 min
60 min
120 min
30 min
180 min
240 min
REST TIME
Indicated by an audible alarm and:
Air Leak LED permanently lit yellow
(when activated) (6).
Blockage LED permanently lit yellow
(when activated) (7).
Service LED permanently lit amber (8)
(NO AUDIBLE alarm).
Audible alarm, no LED lit.
All 'Inflate time LEDs' and 'Rest time
LEDs' flashing (when activated).
10