Toto® Touch Control Unit Operating Guide
1.
Start, stop and pause the control unit.
2.
Air cell side selection - Left only, Right only or Both.
3.
Air cell inflation time - Length of time the air cell remains inflated; minimum 30 minutes, maximum
240 minutes.
4.
Air cell rest time - Length of time the air cells remain deflated and the platform is flat; minimum 30
minutes, maximum 240 minutes.
5.
Alarm mute and interface lock mode activation.
Turning the Toto® Touch Control Unit On
•
Press and hold the start/stop button
•
A single beep confirms activation along with default lit LEDs.
•
Note the system default settings below; these can be changed as appropriate.
o
Air cell side selection - both sides.
o
Air cell inflation time - 120 minutes.
o
Air cell rest time - 120 minutes.
•
The pump automatically inflates the left side of the platform; this can be changed as appropriate.
Air Cell Inflation Selection
•
Press the side selection button
o
Left and right air cell (Default).
o
Right air cell only.
o
Left air cell only.
•
Side selection refers to the patient's side, as standing at the foot end of the bed.
•
A single beep confirms activation along with a lit LED.
Changing the Air Cell Inflation Time
•
Press the inflate time button
o
30 minutes.
o
60 minutes.
o
120 minutes (Default).
o
180 minutes.
o
240 minutes.
•
A single beep confirms activation along with a lit LED.
Changing the Air Cell Rest Time
•
Press the rest time button
o
30 minutes.
o
60 minutes.
o
120 minutes (Default).
o
180 minutes.
o
240 minutes.
•
A single beep confirms activation along with a lit LED.
Once settings have been selected, lock the interface by pressing the interface lock button
by the padlock LED
Therapy Pause Mode
The therapy pause mode temporarily deactivates all control unit functions, returning the platform to a
flat position.
•
Press the start/stop button
•
The currently selected side selection, inflate and rest time LEDs flash when active.
•
For safety, the alarm activates after 10 minutes. To continue in this mode for another 10 minutes, press
the Alarm Pause button
•
To deactivate, press the start/stop button
unit returns to the previously selected settings.
L & R
R
touch
L
for three seconds.
to scroll through the three available options:
to scroll through the five available options:
to scroll through the five available options:
turning on and a double beep (see Interface Lock).
a single beep confirms activation.
, repeat this process until Therapy Pause Mode is no longer required.
INFLATE TIME
240 min
180 min
30 min
120 min
60 min
60 min
120 min
30 min
180 min
240 min
REST TIME
. A single beep confirms deactivation and the control
9
confirmed