System Error Trouble Shooting Guide
Problem
No LEDs lit; no
• Control unit may
audible alarm
not be attached to a
heard.
power source.
• The fuse may need
replacing.
Air-leak alarm
• Pressure has
LED (6) lit
fallen below the
permanently
minimum operational
yellow with an
requirements.
audible alarm.
Blockage alarm
• Airflow obstruction
LED (7) lit
between the control
permanently
unit and the platform.
yellow with an
audible alarm.
Service LED (8)
• Service required.
permanently lit
amber.
Interface panel
• Interface panel is
is lit but
locked.
unresponsive.
An audible
• Loose or absent
alarm is
power lead.
sounding; no
• Mains power
LEDs lit.
accidentally turned
off.
• Mains power failure.
Do not open the control unit. Opening the unit could cause personal injury or equipment damage.
NB. The Toto® Touch control unit completes self diagnostic evaluation on start up.
Cardiopulmonary Resuscitation Emergency Deflation System
The Toto® system features a quick release Cardiopulmonary Resuscitation deflation valve which allows for
rapid deflation for emergency procedures.
In the event of an emergency, firmly pull and release the yellow tab.
To re-inflate the system, replace the Cardiopulmonary Resuscitation tag ensuring both sealing connectors
are firmly attached and restart the Toto® Touch control unit
Cardiopulmonary Resuscitation
Deflation Valve Closed.
Inspection and Care
The Toto® lateral turning system cover, platform assembly, air cells, Cardiopulmonary Resuscitation deflation
valve and tubing set and control unit each have a unique GS1 compliant barcode which should be retained
and used in the event of a warranty claim. NEVER remove these labels.
Cover
Regularly inspect the outer and inner surfaces of the cover for signs of damage.
Report damaged covers to the ward or department manager or appropriate healthcare
professional.
Cause
1.
2.
3.
4.
If the problem persists, contact Frontier Therapeutics
Limited Tel: +44 (0) 330 460 6030 for assistance.
1.
2.
3.
4.
If the problem persists, contact Frontier Therapeutics
Limited Tel: +44 (0) 330 460 6030 for assistance.
1.
2.
3.
If the problem persists, contact Frontier Therapeutics
Limited Tel: +44 (0) 330 460 6030 for assistance.
Please contact Frontier Therapeutics Customer Services
for support Tel: +44 (0) 330 460 6030
The Control Unit continues to function even though the
service LED remains lit.
1.
2.
1.
2.
3.
Initial Checks / Trouble Shooting
Check that the mains power outlet is live and active.
Check the control unit is switched on.
If following points 1 and 2 does not resolve the problem,
switch off and unplug the control unit:
- Check the mains plug fuse (3A).
- Check the control unit fuse (1A).
Restart the control unit.
Press the alarm pause button.
Check the Cardiopulmonary Resuscitation valve is
closed and correct.
Check all air cells and hoses for any air leakage.
Resolve and wait for the alarm to reset.
Press the alarm pause button.
Check for twists or kinks in the air hoses between the
platform and control unit.
Resolve and wait for the alarm to reset.
Check that screen lock LED is lit amber (9).
Press and hold the alarm mute button for four seconds
to deactivate.
If the problem persists, contact Frontier Therapeutics
Limited Tel: +44 (0) 330 460 6030 for assistance.
Reconnect mains power lead to the control unit.
Check the mains power wall switch is turned on.
Attempt to restart the control unit.
If the problem persists, contact Frontier Therapeutics
Limited Tel. +44 (0) 330 460 6030 for assistance.