Policy and declaration by manufacturer - Electromagnetic emission
GP-UH532 is intended to be used in the electromagnetic environment specified below.
Ensure that this product is used in the following environment.
Immunity test
Immunity to conducted dis-
turbances, induced by radio-
frequency fields
IEC 61000-4-6
mmunity to radiated, radio-
frequency, electromagnetic
field
IEC 61000-4-3
Note 1: In 80 MHz and 800 MHz, the higher frequency range will be applied.
Note 2: The above policy may not be applied to all the cases.
Absorption by buildings, objects, and human body, or reflection affects the propagation of electromagnetic power.
*1 The strength of the electromagnetic field of the fixed radio transmitter, including the ones from wireless telephone base sta-
tion, land mobile station, amateur radio, AM/FM radio broadcast, and TV broadcast, cannot be correctly assumed. In order
to evaluate the electromagnetic environment in relation to the fixed radio transmitter, it is necessary to conduct a local
inspection. Check for the normal operation of GP-UH532, if the measured electromagnetic strength for GP-UH532
exceeds the above described compatible level. If any abnormal behavior is observed, relocation of GP-UH532 or other
relieving measures is required.
*2 In the frequency range of 150 kHz to 80 MHz, the field strength shall be less than 3 V/m.
IEC 60601
Test level
3 Vrms
3 Vrms
80 % AM (1 KHz)
80 % AM (1 KHz)
150 kHz to 80 MHz
3 V/m
3 V/m
80 % AM (1 KHz)
80 % AM (1 KHz)
80 MHz to 2.5 GHz
Compatible level
Electromagnetic environment - Policy
Do not use a portable or mobile type wire-
less communication device under the con-
dition where the recommended separation
distance between any component of
GP-UH532 including cables is not ensured.
The distance should be calculated based
on the formula corresponding to each fre-
quency of the device.
Recommended separation distance
d = 1.2
d = 1.2
d = 2.3
P represents the rated maximum output of
the transmission device and watt (W) is
used as a unit. For the separation distance,
d meter (m) is used.
The strength of the electromagnetic field of
the fixed radio frequency device shall be
determined by the local inspection*
ever, it should fall below the compatibility in
each range of radio frequency*
Interference may occur in the vicinity of the
device with the following symbol.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
1
. How-
2
.
5