4. OPERATION PROCEDURES
Operation Start
Refill Valve
Sample inlet
(1) Run water bath, oil bath, or cooling water circulator at desired temperature.
(2) Turn RE unit ON (|)
(3) Rotate Refill Valve to close sample inlet. (The color marking on stopcock faces front)
(4) Sample liquid may be fed as follows. The procedure differs by the way of sample feed.
(4-A) Where samples are continuously fed by
sample feed tube.
① Connect sample inlet and sample container
with sample feed tube.
② Lower slide panel gently to immerse
evaporation flask in the bath.
Do not let the bath fluid overflow.
③ Press and hold the Run/Stop key to start
rotating evaporation flask.
④ Start up the pressure reducing device and
decompress RE unit.
⑤ Rotate Refill Valve slowly so that the color
marking on the stopcock faces down. Sample
liquid is drawn into the flask.
Refill Valve
Feeding sample rapidly may
bump the liquid. Rotate the
stopcock slowly to avoid a loss
of sample, and other
complications.
Color marking
(4-B) Where samples are not continuously fed
(manually added)
① Detach evaporation flask and pour sample
liquid directly in it, then re-attach the flask to
Vapor Tube.
② Start up the pressure reducing device and
evacuate RE unit.
③ Lower slide panel gently to immerse
evaporation flask in the bath.
Do not let the bath fluid overflow.
④ Press and hold the Run/Stop key to start
rotating evaporation flask.
27
Water bath BM312-G
Oil bath BO312-G