Descargar Imprimir esta página

KNF Yamato RE212-G Manual De Instrucciones página 163

Rotavapor

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 258
8. TROUBLESHOOTING
Troubleshooting Guide
Symptom
Unit does not turn on
when the Power
switch is turned ON (|).
Temperature does not
rise when the
Run/Stop key is ON
Possible causes
Power supply failure
Power cable disconnection
Power switch failure
Controller failure
Independent overheat prevention
device is activated
Overcurrent protection fuse
(15 A) is blown
External temperature is below
5 °C
Independent overheat prevention
device is activated
Temperature sensor failure
Controller failure
Heater failure
Power supply failure
Check power supply voltage
Must be 180-250 V AC
Insert power cable firmly deep into the
power inlet. See "Make power
connection properly." (P.11)
Replace relevant parts
Replace relevant parts
Press the reset switch.
See "Independent Overheat Prevention
Device Reset" (P.21)
Replace relevant parts
Operating ambient temperature range
is 5 to 35 °C
Press the reset switch.
See "Independent Overheat Prevention
Device Reset" (P.21)
Replace relevant parts
Replace relevant parts
Replace relevant parts
Check power supply voltage
Must be 180-250 V AC
28
Possible measures

Publicidad

loading