5. HANDLING PRECAUTIONS
Warnings and Cautions
ALWAYS run equipment within specified temperature range.
Never attempt to operate unit outside of specification range. Equipment malfunction or
damage may result.
Overnight and extended storage.
When unit is not in operation during the night or for extended period of time, be sure to
turn OFF (○) the Power switch and disconnect power cable.
Discharge water/oil and clean the reservoir if unit will be in storage for a long period of
time. See "6. MAINTENANCE PROCEDURES" (P.24). Failure to do so may lead to
buildup of scale, or malfunction due to corrosion.
Power loss recovery.
When a power loss occurs during operation and then restored, unit may resume operation
or remain on standby. These actions can be selected through user setting. See "Auto-
resume Function" (P.18) for setting procedure; default setting is "OFF".
Exercise caution when processing heat-generating substances.
Note that temperature reading may not be consistent when processing heat-generating
samples.
Use calibration offset function to correct temperature reading.
If there is a discrepancy between temperature reading and actual liquid temperature, see
"Calibration Offset" (P.17) to make a correction.
Inspect regularly.
Regular inspection and maintenance are highly recommended to ensure proper
operation. See "6. MAINTENANCE PROCEDURES" (P.25) for detailed instructions.