4. OPERATION PROCEDURES
Independent Overheat Prevention Device Reset
Be sure to reset independent overheat prevention device after confirming
that temperature of the bath reservoir has become room temperature.
Unit is hot when the device is activated. Exercise vigilance in order to avoid getting burnt.
Resetting independent overheat prevention device
Independent overheat prevention device shuts off power supply to the controller when bath
temperature has risen beyond the device activation temperature (fixed).
When independent overheat prevention device activates, the display will be blank despite the
Power switch in the ON (|) position.
1 Turn power off
2 Reset independent overheat
prevention device
③
3
Turn power on
④
Contact original dealer of purchase if unit will not restart.
1. Turn the Power switch OFF (○).
①
2. Address the cause of independent overheat
3. Check that unit is cooled to room temperature and
The device cannot be reset while unit is hot. Wait
prevention device activation; low water/oil, ambient
temperature exceeds 36 °C, etc.
press the reset switch of independent overheat
prevention device.
until bath temperature falls below 60 °C.
Turn the Power switch ON (|).
Control panel display will turn on.