Xylem SI Analytics TitroLine 6000 Manual De Instrucciones página 92

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

92
Interchangeable units
Compatibility:
Recognition:
Valve:
Cylinder:
Hoses:
Bracket for supply bottle:
Materials:
Dimensions:
Weight:
Dosing accuracy:
Dosing accuracy of the Titrator TitroLine
Interchangeable.
unit
type No.
WA 05
WA 10
WA 20
WA 50
units are compatible to the titrators TitroLine
Piston Burette TITRONIC
automatically through RFID. Recognition of unit size and characteristics of the
Titration- or dosing solution
volume neutral cone valve made from fluorocarbon polymers (PTFE), TZ 3000
borosilicate glass 3.3 (DURAN
FEP hose set, blue
suitable for square glass bottle and misc. reagent bottles
borosilicate glass DURAN
polypropylene,
15 x 34 x 22.8 cm (W x H x D) incl. reagent bottle
approx. 1.2 kg for interchangeable unit WA incl. empty reagent bottle
after DIN EN ISO 8655, part 3
Accuracy:
Precision:
(Depending on the used interchangeable unit)
®
6000/7000 with WA interchangeable units:
Volume
Tolerances of the Ø i
of the glass cylinder
[ml]
5.00
10.00
20.00
50.00
®
500
®
)
®
, fluorocarbon polymers (PTFE), stainless steel,
0.15 %
0.05 - 0.07 %
[mm]
± 0.005
± 0.005
± 0.005
± 0.005
Status Nov 21 2013
®
6000/7000, TitroLine
Dosage error*
Reproducibility
according to
100 % volume
[%]
± 0.15
± 0.15
± 0.15
± 0.15
®
7000 and
[%]
0.07
0.05
0.05
0.05

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics titroline 7000

Tabla de contenido