Chapter 2 - Unpacking and first Operation
2.4.5

Connection of analytical balances

Analytical balances are to be connected to the RS232-2 using an appropriate cable
®
2.4.6 Connection of SI Analytics ID electrodes to TitroLine
7000
The connector of the ID electrode contains a bead. This bead can be used as a marker for connecting the
electrode to the mV/pH socket. The bead should possibly point upward to the reference socket or in between
(also refer to Fig. 4). The identification of the ID electrode is thereby simplified. Data of the connected ID
electrode are read out immediately after the connection and stored in the titrator. This includes the calibrating
data, such as zero point and slope, date of the calibration, buffer solutions used, the serial number and type of
electrode.

2.5 Setting the Language of the Country

The ex-factory default language setting is English. When the piston burette is switched on, the main menu will
appear once the boot sequence is completed:
Fig. 5
Using <SYS/<F7> or <MODE>, followed by <System settings> you navigate to the system settings. The very first
menu is to be used for setting the language of the country:
Fig. 6
Use <ENTER>/<OK> to call the menu. Select the national language using the <↑↓> arrow keys, confirm it with
<ENTER>/<OK>: