Specifications Titrator Titroline ® 6000/7000 - Xylem SI Analytics TitroLine 6000 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

90
1.2 Specifications Titrator TitroLine
CE sign:
ETL sign:
Country of origin:
The following solvents/titration reagents are allowed to be used:
− All common titration solutions.
− As reagent water and all non-aggressive non-organic and organic fluids are allowed. If using combustible
fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the chemical industry.
− For fluids with higher viscosity (≥ 5 mm
speed can be adjusted.
− Fluids with viscosity over 20mm
Measuring input:
Measurement range
pH
pH
mV
U [mV]
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
®
TitroLine
6000/7000 titrator to "warm up".
Measurement input:
T [°C]
Measurement input:
* The measurement uncertainty of the sensor probe has to be taken into account as well
EMC compatibility according to the Council Directive: 2004/108/EG;
applied harmonized standards: EN 61326-1:2006
Low-voltage directive according to the Council Directive 2006/95/EG
Testing basis EN 61 010, Part 1
Conforms to ANSI/ UL Std. IEC 61010-1
Certified to CAN/ CSA Std. C22.2 No. 61010-1
Germany, Made in Germany
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
pH/mV-input with 24 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or
additional with BNC socket insert (Z 860)
Reference electrode 1 x 4 mm socket
®
TitroLine
7000 only:
®
TitroLine
7000 only
- 4,0 ... 18,00
- 2000 ... 2000
- 2020 ... 2020
Temperature probe - connector for a Pt 1000 resistance thermometer
Connection: 2 x 4 mm - sockets and 1 x 2 mm socket.
Measurement
range
- 75 ... 175
Karl Fischer (Dead-stop) connector for double platinum electrode
Polarisation voltage variably adjustable from 40 ... 220 mV.
Connector: 2 x 4 mm - sockets.
Measurement
range
0 ... 100
I [µA]
®
6000/7000
adjustable damping settings of the pH/mV measuring signal
RFID receiver for SI Analytics ID electrodes
Over range
Display
resolution
- 3,1 ... 18,00
0,001
0,1
Display resolution
0,1
Display resolution
0,1
State Nov 21 2013
Measurement
accuracy* without
sensor probe
0,002 ± 1 Digit
0,10 ± 1 Digit
Measurement
accuracy* without
sensor probe
0,2 K ± 1 Digit
Measurement
accuracy* without
sensor probe
0,2 ± 1 Digit
Input
resistance
[Ω]
> 1 • 10 13
> 1 • 10 13

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics titroline 7000

Tabla de contenido