Technical Specifications of the Titrator TitroLine® 6000/7000

1.1 Summary

®
The TitroLine
6000/7000 is a potentiometric titrator and suitable for the following applications:
The possible range of titrations includes pH, mV, µA with a maximum of 50 (TitroLine
memorisable methods.
− The examples of possible use of the TitroLine
− Acid and base determination in aqueous solutions such as p and m value, titration of strong and weak acids
and bases
− Redox titrations such as iodometry, manganometry, chromatometry, and COD determinations
− Other mV titrations, e.g. chloride
− Titrations using ion-selective electrodes, e.g. calcium, fluoride, copper, lead ions
− Indices such as OH number, iodine number, and saponification number.
Applications which are only feasible with the TitroLine® 7000:
Read-out and store calibration data from SI Analytics ID Electrodes.
Titration up to two inflection points examined as the titration of calcium and magnesium
pH-stat titrations
Non-aqueous potentiometric titrations such as TAN and TBN
Pre-dosing with a connected piston burette
Connection and use of an autosampler TW alpha plus/TW 7400
Compatibility with TitriSoft starting at version 3.0
These methods are mere examples; further applications can be found in food technology, pharmaceuticals,
biochemistry, photofinishing, environment, quality control, and process monitoring.
In addition, the TitroLine
− Manual titrations with or without calculation of the result
− Dosing
− Preparation of solutions
Each method allows for the setting of a variety of dosing and filling rates.
Solutions to be used:
Virtually, any liquids and solutions with a viscosity of < = 10 mm² / s such as concentrated sulphuric acid may
be used. However, one has to avoid the use of chemicals that may attack glass, PTFE or FEP or that are
explosive, such as hydrofluoric acid, sodium azide or bromine! Suspensions containing high solids
percentages may clog or even damage the dosing system.
!
General provisions:
The safety guidelines that are applicable to the handling of chemicals have to be observed under all
circumstances. This applies in particular to inflammable and/or etching liquids.
Guarantee
We provide guarantee for the device described for two years from the date of purchase. This
guarantee covers manufacturing faults being discovered within the mentioned period of two years.
Claim under guarantee covers only the restoration of functionality, not any further claim for damages
or financial loss.
Improper handling/use or illegitimate opening of the device results in loss of the guarantee rights.
The guarantee does not cover wear parts, as lobes, cylinders, valves and pipes including the thread
connections and the titration tips. The breach of glass parts is also excluded. To ascertain the
guarantee liability, please return the instrument and proof of purchase together with the date of
purchase freight paid or prepaid.
®
6000/7000 comes with the functionalities of the TITRONIC
!
®
6000/7000 include:
®