Fig. 138

7.3 Connection of Printers

The results, calibration data and methods can be printed on the following media:
HP PCL compatible printer (A4)
Seiko DPU S445 (Thermo paper 112 mm width)
On the USB stick in PDF- and CSV-format
To connect the printers to the burette please use the USB socket. When printing, please check whether the
correct printer is connected. It is not possible to print „HP"printer layouts on another thermal printer or vice versa.
The printer settings should always be checked and adjusted after changing the printer.
Fig. 139
Only one printer should be connected for one Titrator because an automatic printer recognition is not activated.
Print PDF is the default setting. If you select "Print PDF", please make sure that a USB stick is connected to the
device.
7.4
Connection of Autosampler (only TitroLine

7.4.1 Connection of sample changer TW alpha plus

The sample changer TW alpha plus is connected to the RS232-2 (RS2) of the titrator with cable TZ 3087. The
settings of the RS232-2 interface must be changed to 4800, No, 7.2:
Chapter 7 – Connection of Analytical Balances and Printers
®