Xylem SI Analytics TitroLine 6000 Manual De Instrucciones página 139

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Drift-controlled acceptance of the measured value in terms of mV/min is set by selecting "normal", "fast or "user-
defined". The drift values at predefined in terms of in mV/min for normal and fast drift:
Normal drift
20 mV/min
Fast drift
50 mV/min
Small drift value = slow and precise
Large drift value = fast and "less precise"
The following parameter selection can be made for user-defined drift setting:
Minimum holding time [s]:
Maximum holding time [s]:
Measuring time t: [s]
Drift [mv/min]
Fig. 87
If normal or fast drift was selected before, the values will be defaulted for user-defined drift. In the present case,
for instance, 20 mV for normal drift:
Fig. 88
Drift-controlled acceptance of the measured value is used in most applications. However, there are applications
in which the setting of a fixed holding time for measured value acceptance following the titration step is
recommendable. Examples hereof include titrations in non-aqueous media. In the case of dead-stop titration no
holding time other than the fixed one can be selected. The fixed delay time can be set between 0 and 999
seconds:
Chapter 4 Method Parameters
01 - 99
01 - 99
01 - 99
01 - 99
139

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics titroline 7000

Tabla de contenido