Figure 7�1 – LithoVue™ Flexscope
Articulation Lever
Connector
Cable and Plug
Strain Relief
b. Advance the guidewire slowly, observing the live video image for initial entry of the guidewire
into the field of view.
c. Stop advancing the guidewire when the stiff portion of the guidewire emerges from the distal
tip and is visible on the live video image.
d. Use fluoroscopy or other imaging modality to confirm that the LithoVue Flexscope's shaft
has straightened.
e. Slowly withdraw the LithoVue Flexscope from the patient. Monitor the withdrawal using spot
fluoroscopic imaging or other imaging modality. If you feel resistance, investigate the source
of resistance and take remedial action before continuing to withdraw the LithoVue Flexscope.
f. Dispose of the LithoVue Flexscope in accordance with hospital, administrative and/or local
government policy.
38
Black (K) ∆E ≤5.0
Articulation Section
Handle
Accessory Access Port
Irrigation Port
Flexible Shaft
Distal Tip
User's Manual